Stability of aciclovir in elastomeric infusion devices
Abstract:
Objectives: To investigate the stability of aciclovir solutions in elastomeric devices used for outpatient parenteral antimicrobial therapy (OPAT).
Methods: Triplicates of two elastomeric devices, Accufuser and Easypump II, were filled with a solution of 200 mg, 2400 mg, and 4500 mg aciclovir in 240 mL 0.9% w/v saline. Devices were stored at room temperature for 14 days, followed by 24 hours storage at 32°C. Assessment using a stability indicating assay, pH and subvisible particle analysis was undertaken at 11 time points throughout the study.
Results: Aciclovir solution at 200 mg and 2400 mg in 240 mL was stable for 14 days at room temperature (<20°C) and 24 hours of 32°C 'in-use' temperature exposure, remaining above the 95% limit for NHS stability protocols. The high dose was also stable for 14 days at room temperature, but when stored at 32°C there was precipitation of aciclovir within 4 hours in both devices. The precipitate was confirmed as aciclovir and precipitation was not a sign of chemical degradation.
Conclusions: Aciclovir concentrations above 2400 mg/240 mL are liable to precipitation and cannot be recommended for OPAT services because of heightened risks of nephrotoxicity. Aciclovir solution can be given as a continuous 24-hour infusion for OPAT services at a concentration range of 200-2400 mg in 240 mL in Accufuser and Easypump II elastomeric devices following 14 days storage at room temperature, protected from light.
Reference:
Sime FB, Wallis S, Jamieson C, Hills T, Gilchrist M, Santillo M, Seaton RA, Drummond F, Roberts J; BSAC OPAT Drug Stability Testing Programme. Evaluation of the stability of aciclovir in elastomeric infusion devices used for outpatient parenteral antimicrobial therapy. Eur J Hosp Pharm. 2023 Dec 6:ejhpharm-2023-003784. doi: 10.1136/ejhpharm-2023-003784. Epub ahead of print. PMID: 38071555.